DPM-6000 PM Monitor PM2.5/PM10
Cantidad : 10,000 Precio : 0-50,000
Overview Of DPM-6000 pm monitor Inhalable particle content is one of the most important indexes for measuring air quality. Combining years of experience in developing environment and safety monitoring instruments as well as the latest demand at home and abroad, our company has launched a new generation DPM-6000 atmospheric particulate monitor with high accuracy, reliability, stability and economy. Our pm monitor can measure PM10, PM2.5 and dust particles of other cut sizes, widely applied in inhalable particles concentration detection in environmental atmosphere monitoring.
Design of instrument meets US EPA and national standard
Adopt PM filter membrane enrichment technology
Particle cutter meets EPA standard TSP, PM10, PM2.5
Use simple, reliable mechanical transmission structure and modular design to ensure the measurement is accurate and reliable, and the pm monitor can operate stably for a long term.
PrincipleFeaturesSpecificationApplication
Principle Of DPM-6000 PM Monitor PM2.5/PM10
The monitoring principle of the pm monitor is mainly based on enrichment technology of atmospheric particulate and beta ray attenuation method. The pm 2.5 monitoring equipment concentrates on particulate matters whose aerodynamic equivalent diameter is smaller than 2.5/10μm on the rolled filter membrane. The air pollution analyzer adopts low-radiation 14C as a radioactive source. The emitted Beta-ray passes through the filter membrane with contaminants and will be received by the detector. The pm 2.5 measurement device is composed of a photo-multiplier tube (or Geiger detector), plastic scintillator and counter. The thickness of enrichment can be analyzed quantificationally according to Beta-ray energy that detector received, then the PM2.5/10 concentration in the air can be calculated.
Features Of DPM-6000 PM Monitor PM2.5/PM10
The design of the gas path keeps filter paper, a radioactive source and detector immovable during enrichment and measurement, ensuring a good lower detection limit.
US EPA standard beta ray attenuation method, which spares the complicated operations of TEOM, such as manually replace filter paper and weight. Compared to the light scattering method, our pm monitor has higher measurement accuracy.
Cutter, based on simulation verification of aerodynamics, will provide a various selection of accurate particle.
Sizes: PM1, PM2.5, PM10, TSP.
Sampling flow is measured by a mass flow sensor and corrected with collected ambient temperature and pressure to get the sampling flow of the field.
Use C14 source which is of low-density, low activity, and long half-life to realize stable measurement. No special protection for the pm measuring device is needed. Reliable and safe.
Flexible paper-pressing structure. Mechanical driving design of the pm 2.5 monitoring equipment is simple and reliable; greatly reduce paper breaking and paper jam and avoid the error caused by the moving of filter paper.
The pipeline heating system is used to eliminate interference from ambient humidity and make the detector fit into sudden.
Flow control system with precise closed-loop feedback; stable sampling flow and small error.
Various external ports, such as RS232/RS485, 4~20mA and etc.
Low maintenance frequency of the pm measuring device, once a year.
Specification Of DPM-6000 PM Monitor PM2.5/PM10
Measuring principle |
Beta ray attenuation method |
|
Performance Index |
Measuring Range |
0~1000μg/m3³or 0~10000μg/m3 (optional) |
Radioactive Source |
14C (carbon-14), radioactivity≤60μCi |
|
Accuracy |
≤5% |
|
Display Resolution |
1μg/m3 |
|
Lower Detection Limit |
5μg/m3 |
|
Flow Error |
≤±2% (generally set at 16.67L/min) |
|
Sampling Flow Stability |
≤±2% working point flow/24h |
|
Sampling Period |
≤1h (time is settable) |
|
Sampling Flow |
Default 16.67L/min |
|
Parallelism |
PM10≤10%, PM2.5≤15% |
|
Filter Paper Tape |
Filter Paper Tape |
Glass fiber, inner diameter: 40 mm, width:30 mm, service time 1 month (work period 1 h) |
Interface Para. |
I/O |
Rs232, RS485, 4-20mA, GPRS (optional) |
Data Storage |
One year |
|
Working Condition |
Cutting header |
Ambient temperature (-40~50)℃, ambient pressure (80~106kPa) |
Detector |
Ambient temperature (5-40)℃, ambient pressure (80-106kPa), humidity (≤90%) |
|
Power Supply |
AC (220±22)V, (50±1)Hz |
|
Dimension & Weight |
483×407×322(mm) (L×W×H); Monitor 30kg, Pump 8kg |
Application Of DPM-6000 PM Monitor PM2.5/PM10
Inhalable particle real-time monitoring in industry,, construction, residential district and etc.
Environment assessment, permission, and pollution forecast and pre-warning.
Ambient air quality monitoring and mobile air monitoring experimental station.
Occupational disease research, and long-term background environment research.
Secondary emission monitoring (coal storage area, wharf, etc.).
HANGZHOU ZETIAN TECHNOLOGY CO., LTD is a professional air pollution monitoring instruments manufacturers, we provide Water Quality Analyzers, Continuous Emission Monitoring & Process Control, Gas Analysis Modules, air quality monitor and etc. Want to know more? Please contact us.
Persona a contactar : Alex Wang, +0086 18957111635
Buen trato: comprar del vendedor
Por favor, lea nuestras condiciones de uso. También puede visitar nuestras preguntas frecuentes y ver nuestra información sobre los riesgos relacionados con la falsificación.
Esta página es acerca de los importadores y exportadores de DPM-6000 PM Monitor PM2.5/PM10 Buscar en la categoria : Informática Buscar en la categoria : monitor, pm10 |
Miércoles 14 septiembre 2016
Cantidad : 5000 - Precio : 50,00 €
Bellows Seal Stop Valves(Straight-Through Type) DIN Bellows Seal Stop Valve DIN Bellows Globe Valve DIN Faltenbalgventil Durchgangsform DIN Válvulas de fuelle DIN ventil uzavírací s vlnovcem přímý DIN Standard seteventil med belg (Rett løp modell) DIN Standardında Metal Körüklü...
ETM Armaturen
- 59590 - Henri Durre
- 86 577 67953935
Miércoles 14 septiembre 2016
Cantidad : 5000 - Precio : 50,00 €
Gland Packing Globe Valves ( Straight- Through Type) DIN Standard Gland Packing Globe Valve DIN Globe valve DIN Stop valve DIN Absperrventil Durchgangsform DIN Válvulas de globo DIN ventil uzavírací s klasickou ucpávkou přímý DIN Standard seteventiler med pakkboks (rett løp...)
ETM Armaturen
- 59590 - Henri Durre
- 86 577 67953935
Miércoles 14 septiembre 2016
Cantidad : 5000 - Precio : 50,00 €
Long life-Heavy Duty Bellows Seal Stop Valves DIN Standard Long Life-Heavy Duty Bellows Sealed Stop Valves DIN Long life bellows Seal Stop Valve DIN Long life bellows globe valve DIN Long bonnet bellows globe valve DIN Faltenbalgventil lange Bauform DIN Válvulas de fuelle de...
ETM Armaturen
- 59590 - Henri Durre
- 86 577 67953935
![Pago seguro](https://es.algomtl.com/images/side_paiement.jpg)