Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Persona a contactar : Bruce Lu, +86 29 89284429
Buen trato: comprar del vendedor
Por favor, lea nuestras condiciones de uso. También puede visitar nuestras preguntas frecuentes y ver nuestra información sobre los riesgos relacionados con la falsificación.
Esta página es acerca de los importadores y exportadores de Normal Problem and Solution in GC Test Buscar en la categoria : Varios / Descuento Buscar en la categoria : problem, normal, test, solution |
Martes 04 junio 2013
VOTRE COURRIER POSTAL SUR INTERNET (BACKOFFICE) www.ubidoca.com Vous voyagez très souvent ? Vous travaillez beaucoup à l'étranger ? Vous avez plusieurs domiciles ? Nous fournissons la solution la plus efficace pour gérer vos affaires et votre courrier postal à distance sur Internet. Nos services - Mise de votre courrier postal sur Internet dès réception par notre équipe. - Vous pouvez le consulter 24h/24, o...
UBIDOCA SARL
- courrier
- 01220 - Divonne les Bains
- 04 50 20 49 70
- 06 18 63 06 53
Miércoles 20 mayo 2015
AI101P Single Port KVM over IP The KVM over IP technology (IP-KVM) combines digital remote KVM access via IP networks with comprehensive and integrated system management. The IP-KVM defines a new class of remote KVM access devices that can save your money, time, space, and equipment...
ACAFA Information Co.
- epanio
- 335 - Dasi Township
- +88 6 33 07 13 00
Domingo 29 marzo 2015
Cantidad : 100 MT FAS - Precio : Negotiation
Myanmar Rice Pawsan/NgaKywe/Emata/Zeera Pawsan/ Ngakywe-1006-3000-010 Pawsan/ Ngakywe Super 100% 1 Grain Composition (a) Whole Kernels & Head Rice 95% 2 (b) Big Brokens 5% 3 N.B Size of Big Brokens (-) 0.75 (+) 0.50 4 Paddy 15 grains per kg 5 (b) Foreign Matters - 6 Maximum...
win-enterprise
- 11081 - yangon
- 9515536846
![Pago seguro](https://es.algomtl.com/images/side_paiement.jpg)