Normal Problem and Solution in GC Test
More than 90% of gas chromatographic problems occur in the inlet, and it is necessary to see which problems may occur in the inlet.
Problem No.1: Leak
If there is a leak, the pressure on the inlet may not be maintained, and the instrument will alarm. If the problem cannot be solved in time, the instrument will automatically turn off the pressure. Leakage is two-way. The carrier gas will leak out and the air will leak in. The air entering the chromatographic system, especially oxygen, is highly destructive, especially at high temperatures, where oxygen in the system can damage the stationary phase coating of the column. It results in much shorter of column life.
The main leak points as following:
1. Septum; the holes in the septum will become larger and larger after repeated use, and the seal will be problematic. For gas chromatography running around the clock, it is better to replace one septum per day.
2. O-ring outside the liner; long-term heating and cooling will also lead to rubber aging, sealing is reduced, O-rings need to be replaced in time.
3. Column seal; After repeated tightening, the sealing performance will also be reduced, try not to reuse it as many times as possible.
Problem No.2: Block
In general, the carrier gas and gas samples will not block the pipeline. Only when the sample contains some high-boiling compounds, some samples will condense in the split flow path due to untimely discharge, resulting in blockage of the pipeline or Partial blockage, which directly affects the accuracy and stability of the split ratio. The chromatographic results is that the repeatability of peak area is not good.
Problem No.3: Sample Residue
Although we have purged through the septum and shunted to minimize sample residue, there are many stubborn samples where the sample remains, including septa, liners, shunt plates, and even the front of the column. Replacing these consumables on the inlet can effectively solves the problem of sample residue. Selecting more inert liner and column also avoids sample residue better.
Persona a contactar : Bruce Lu, +86 29 89284429
Buen trato: comprar del vendedor
Por favor, lea nuestras condiciones de uso. También puede visitar nuestras preguntas frecuentes y ver nuestra información sobre los riesgos relacionados con la falsificación.
|
Esta página es acerca de los importadores y exportadores de Normal Problem and Solution in GC Test Buscar en la categoria : Varios / Descuento Buscar en la categoria : normal, test, problem, solution |
Thursday 14 March 2013
Cantidad : 1 - Precio : 400,00 €
Anilutum Brand Autumn England, Men's long-sleeved shirt easy care casual S219133 Made In China Model Number S243406 Brand Name Anilutum Gender Men's Material Polyester Season Winter Type Suit Brand: Anilutum Style: Fashion Type: Men's Suit Online: 9AM-5PM Aoniludan...
H&Y International Fashion Limited
- anilutum
- EC1V 2NX - London
- +86 13724036041
Tuesday 08 September 2015
Module parameters HD-720P-38*38mm 1.0megapixel low-illumination HD web camera Network Camera - Sensor type 1/4 inch、1.0megapixel , CMOS low-illumination width dynamic sensor - Minimum illumination Color: 1 Lux/F1.2, black and white: 0.02Lux/F1.2 - Electronic shutter 1/5 second ~ 1/50...
ZT Technology Group Ltd.
- ellazhan
- 852 - Fo Tan
- +85 2 26 75 75 58
Tuesday 28 April 2026
WT- 480T / R WT-480T with hidden electric condenser microphone or you can use lapel microphone. WT-480R with hidden super thin speaker, could be applied without earphone conditions. Match with dual slots charging stand HDC-202 or 30 slots portable charging HDC-30, easy to recharge battery...
Meicheng Audio Video Co., Ltd.
- meicheng
- 23511 - New Taipei City
- 886 2 82280311