CTnI test kit (FIA) CE Marked
Product Name
Troponin I (cTnI) Test Kit (Immunofluorescence Chromatography)
Purpose
To detect the cTnI in human plasma/serum/whole blood in vitro quantitatively.For Professional Use only.
Principles of Detection
The cTnI Test is a fluorescent lateral flow immunoassay. The test consists of: 1) a conjugate pad containing cTnI antibody conjugated with fluorescent microspheres and a control antibody (goat anti-chicken IgY antibody) conjugated with fluorescence microspheres, 2) a nitrocellulose membrane strip containing a test line (T line) and a control line (C line). The T line is pre-coated with another cTnI antibody, and the C line is pre-coated with a control line antibody (chicken IgY antibody).
cTnI concentration in specimen is positively correlated with the fluorescent density of the microspheres, after the measure and calculation of the quantitative fluorescence immunoassay analyzer, then the analyzer would calculate the concentration of cTnI according to the preset calibration curve in the machine and display the result, detection unit is ng/mL. C line should appear in any cases, which indicates the reaction system is effective.
Storage conditions and shelf life
Store in a sealed aluminum foil bag at 2~30°C. After unpacking the aluminum foil bag under temperature 20~30 ºC , humidity 35~65%, please use it immediately.
Shelf life 18 months.
Applicable instrument
Avid fluorescent immunoanalyzer model: AI-IF 600
Reference range
The 95th percentile of the serum and plasma samples of 200 normal persons is <0.3ng/mL.;<br />
The reference value is verified by referring to similar products on the market. Due to ethnic and regional differences, each laboratory can establish its own reference interval according to the actual situation.
Performance Characteristics
1. Linear range
In the range of 0.1ng/mL ~40ng/mL, the r value should be ≥0.990.
2. Accuracy
Detected with the enterprise internal control accuracy reference (C1, C2), the test results should be within the following ranges.
(1) Detection with the enterprise internal control accuracy reference C1, the instrument measurement range is 0.5ng/mL ±15% at 15 minutes.
(2) Detection with the enterprise internal control accuracy reference C2, the instrument measurement range is 5ng/mL ±15% at 15 minutes..
3. Intra-batch precision
Using the enterprise internal control precision reference (C9, C10) each parallel test 10 times, at 15 minutes instrumentation, the concentration value coefficient of variation CV≤15%.
4.Inter-batch precision
For three batches of reagent kits, 10 times of parallel testing with each enterprise internal control precision reference (C9, C10), and the instrument measurement at 15 minutes, the concentration coefficient of variation CV≤15%.
5.Detection limit
Detection with the enterprise internal control of the detection limit reference L1, instrumentation at 15 minutes, the results should be ≤0.1ng/mL.
6.Specificity
Detection with the enterprise internal control specificity reference (C3, C4, C5), the instrument measurement value at 15 minutes should be less than 0.1ng/mL.
Persona a contactar : Yang Shawn, +8618543132823
Buen trato: comprar del vendedor
Por favor, lea nuestras condiciones de uso. También puede visitar nuestras preguntas frecuentes y ver nuestra información sobre los riesgos relacionados con la falsificación.
|
Esta página es acerca de los importadores y exportadores de CTnI test kit (FIA) CE Marked Buscar en la categoria : Medicina Buscar en la categoria : ctni, marked, test |
Thursday 02 April 2015
SPECIFICATION : Frame Spark Carbon / IMP technology / HMX / BB92 / Carbon swingarm with 160PM U-Mono Link / Tapered Headtube IDS SL dropouts for 142 × 12mm BB height adj. Travel Front 120mm Rear 120 - 85 - Climb / Shock 190 × 50 Fork FOX 32 Float Factory CTD Air / Kashima...
INDOBIKESPORT
- 20152 - Medan - Sumatera Utara
- +628994498351
Tuesday 25 November 2025
Cantidad : MOQ200 - Precio : To be discussed
Key Specification: - Color: Various color can be chosed - Input Voltage: 12V DC - Power (W): 8.5W - Noise Level: 47dB - Display Mode:LED - Control Mode: Touch screen - CADR: 40m³/H - Purifying Rate: 99.7% - Length Of Power Cord: 2.2M standard or as per customers'...
Rooman Electrical Appliances Co., Ltd.
- Rooman1
- 523929 - Dongguan
- +13 6 62 88 32 70
Tuesday 25 November 2025
Two way LCD car alarm system with engine start module Detecting method Rechargeable Li-battery Window roll up control Long range monitoring www.autosmartelec.com Ubicación : 1015 central, 519015 Hong Kong, Persona a contactar : Helen Loa, 867563885632
Autosmart Electronic Co., Limited
- 519015 - Hong Kong
- 867563885632